Viscotek - Setting the Standard for GPC
You are here: Skip Navigation LinksHome > Theory > GPC:Light Scattering log in
INSIDE VISCOTEK
"The service provided by Viscotek representatives is unmatched by any other in the industry... A wonderful addition to the polymer community."
 - E.F, Ph.D., Academic Institution
 
Tetra Detector Array
A revolutionary, integrated multiple-detector device designed for the characterization of natural and synthetic polymers and copolymers, proteins, protein conjugates, excipients and other macromolecules. Comprised of modular Refractive Index, Ultra-Violet, Low Angle Light Scattering and Viscometer detectors.
 
HTGPC
The Viscotek High Temperature GPC (HT-GPC) System is a revolutionary advanced detector system specifically designed for the characterization of polyolefins, natural and synthetic polymers, nanoparticles and other large molecules.
 
Nanoparticle and Protein Sizing
Viscotek is proud to introduce its range of Dynamic Light Scattering Detectors featuring rapid, accurate and sensitive sizing for proteins, biomolecules, nanoparticles & polymers and a pair of outstanding technologies unique to Viscotek.
Light Scattering Detector Theory
GPC Theory: Light Scattering Detector
Design
Low Angle Light Scattering Schematic

Theory


The fundamental equation for the scattering of light from polymer solutions is the Zimm equation.
Zimm Equation

M is the molecular weight of the polymer sample and C is the sample concentration. A2 is the second virial coefficient of the solution, which corrects for the interaction of polymer molecules with each other. A2 may be calculated from the concentration dependence of the light scattering signal. Rθ is the excess Rayleigh scattering ratio of the solution above that of the pure solvent, measured at angle θ with respect to the incident beam.
Rayleigh Ratio

I0 is the irradiance of the incident laser. Iθ is the excess intensity of the scattered light above that of the pure solvent at angle θ; in the present case it is the baseline-corrected LALS signal. k is an instrument constant related to the scattered light collection efficiency. A schematic of the optical arrangement of the LALS is shown in Figure 1. Pθ is the particle scattering factor. It is a measure of the angular dissymmetry of the scattered light and is related to the size and the angle at which the scattering is determined. Much of light scattering science is devoted to the determination of Pθ but the beauty of LALS is that it can be ignored.
  • Pθ is exactly equal to unity for all molecules when θ is zero.
  • Pθ is 0.98 for molecules with radius of gyration (RG) of 150 nm when θ is 7 degrees.
  • RG of 150 nm constitutes the upper limit of separation of GPC.
Therefore, with LALS, multiple angle measurements are unnecessary because extrapolation or correction for angular dissymmetry is unnecessary. This reduction to a single angle greatly simplifies the processing of data from multiple detectors. K is a composite of optical and fundamental constants.
Optical and Fundemental COnstants

  • n0 is the refractive index of the solvent. 
  • ν is the refractive index increment of the polymer solution. 
  • NA is Avogadro’s number. 
  • λ0 is the wavelength of the incident light in vacuum. 
  • p is an integer equal to 2 for vertically polarised incident light, 1 for un-polarised.
The latter three parameters (NA, λ0 and p) are instrument constants and can be merged with the detector constants k and I0 in equation 2 to form a new constant which we will call Lals.Cal. Equation 1 can now be rearranged to separate the variables of molecular weight and concentration from the other parameters.
molecular weight and concentration

At normal GPC/SEC concentrations the second virial coefficient A2 is typically insignificant compared to the first term in the denominator, so equation 4 can be simplified by neglecting the A2 term.
molecular weight and concentration Simple

Equation 4 is the general equation but equation 5 will be used in the present discussion for purposes of simplification.

Elution Profiles

The elution profile consists of successive fractions of the eluent sampled at equally spaced time intervals, i. Each fraction will be characterised by its molecular weight Mi and concentration Ci. The term Ci is determined from the following equation, derived in The Deflection Refractometer.
Concentration RI

RIi is the signal from the RI detector at interval i and RI.Cal is the detector calibration constant. Substituting equation 6 into equation 5 yields the elution profile of molecular weight.
elution profile of molecular weight

Note that the array index for the LALS signal is offset from that of the RI detector by an amount δ. This detector offset reflects the fact that the two detectors do not measure each fraction simultaneously. They have a time (volume) separation between them corresponding to the volumes of the two detectors plus the volume of the interconnecting tubing. Equation 7 reveals that molecular weight elution profile is proportional to the ratio of the LALS detector signal to the RI detector signal.

Log M elution Profile

An elution profile for molecular weight on a broad distribution polystyrene sample is shown in Figure 2, overlaid with the RI and LALS signals. Notice that the Mi profile has more noise at each end than in the middle. This is due to the RI and LALS signals being lower in magnitude near the ends of the peaks, so the noise is relatively higher. In fact, the calculation of Mi becomes so unreliable on the ends that the calculation must be truncated at some point and Mi is obtained for the rest of the distribution by extrapolation. The extrapolation is shown in Figure 2 as the dashed line and is much more extensive on the low molecular weight end.

Determination of instrument constants (calibration)

The instrument constants RI.Cal, Lals.Cal and δ are determined from the chromatograms of a narrow distribution polymer standard — for example, polystyrene standard 90K, shown in Figure 3. The offset δis easily determined as the difference in the peak positions. Figure 4 shows the same chromatograms after the offset is applied. The RI calibration constant is determined from the RI peak area and dn/dc as shown previously in Part 1. The LALS calibration constant is determined from the LALS peak area, which is directly proportional to the weight-average molecular weight as shown in the next section.

Narrow Distribution

Narrow Distribution with Detector Offset


Molecular Weight Distribution

The molecular weight distribution can be represented in several ways — the most important being the number-average molecular weight MN and the weightaverage molecular weight MW. MN is defined as the average molecular weight (molar mass) over the successive fractions of the sample, with the statistical weight of each fraction being the number of molecules, or molar concentration Ni. The molar concentration is simply the ratio of the weight concentration Ci and molar mass Mi, which are determined by equations 6 and 7, respectively.
Molar Concentration

The number-average molecular weight is therefore defined as follows:
Number Average

The weight-average molecular weight is defined as the average molecular weight (molar mass) over the successive fractions of the sample, with the statistical weight of each fraction being the mass of molecules, or weight concentration, Ci.
Weight Average

The summations in equations 9 and 10 are directly related to peak areas in the GPC chromatogram. The sum of Ci is proportional to the area of the RI detector peak, per equation 6.


The sum of CiMi is proportional to the area of the LALS detector peak, per equation 5.
LALS Peak Area

Combining equations 10–12 reveals a very simple relationship between MW and the chromatographic peak areas.
Simple Mw

The sum of Ci/Mi is proportional to the area of the molar concentration peak, per equation 8. So MN can also be considered a ratio of peak areas as follows:
M sub N

MW MN

However, molar concentration is a derived function, not a detector signal, so MN does not have the same type of simple relationship to detector peak areas as does MW. The derivation of molar concentration often results in greater error for determination of MN than for MW. The reason for this is illustrated in Figure 5 where the relevant peaks are overlaid for a broad distribution sample, polystyrene in THF. MW can be determined very precisely because it is simply proportional to the ratio of the LALS and RI peak areas, both of which have excellent signal/noise. MN is determined by the peak area of molar concentration, which has considerable noise on the long elution side of the peak. This noise is inherent in a broad distribution sample and arises from the fact that molar concentrations are higher where the molar mass (and therefore the light scattering signal) is lower. MN will therefore be determined by light scattering with less precision than MW where the distribution is broad. Narrow distribution samples will not exhibit this problem.

In summary, the LALS detector provides sensitive, simple and accurate molecular weight measurement for GPC/SEC. By measuring directly at very low angle it avoids the complexity, assumptions and errors inherent in angular correction.
Complete Gel Permeation Chromatography (GPC) / Size Exclusion Chromatography (SEC) systems for the characterization of natural and synthetic polymers, proteins, conjugates and excipients. Integrated Triple and Tetra Detector Arrays (TDAs) featuring Low Angle Light Scattering (LALS) detectors, four capillary differential viscometer detectors, refractometers and UV/VIS detectors; in a single experiment, obtain absolute molecular weight, molecular size, intrinsic viscosity, conformation, branching, aggregation and copolymer composition. Flow Injection Polymer Analysis (FIPA) for fast, accurate and precise average molecular weight, size and intrinsic viscosity in routine QA/QC and process control. Automated Dilute Solution Viscosity (DSV) systems provide relative, inherent, and intrinsic viscosity.